80PK-25
SureGrip
TM
Piercing
Temperature Probe
Instruction Sheet
XWWarning
To avoid electrical shock, do not use this
probe when voltages exceeding 24 V ac rms
or 60 V dc are present. The probe tip is
electronically connected to the output
terminals.
Introduction
The 80PK-25 SureGripTM Piercing Temperature Probe is
designed for use in liquids, gels, soft, and semi-hard
materials. The probe is made of 316 stainless steel, which
is suitable for food service use. The thermocouple junction
of the probe is protected from tip to handle by the stainless
steel sheath. The 40-inch (1-meter) cable is terminated
with a Type K miniature thermocouple connector with
0.792-mm (.312-inch) pin spacing. The 80PK-25 can be
used with any temperature-measuring instrument that is
designed to accept Type K thermocouples and has a
miniature connector input.
Specifications
Type
K,
Standard grade Ni-Cr vs Ni-Al (Chromel vs Alumel)
Measurement Range of Stainless Steel Probe
-40 °C to 350 °C (-40 °F to 662 °F)
Accuracy
(With respect to ANSI MC96.1 Standard Limits of Error)
Note
All error calculations should be done in °C, then
scaled to °F.
Range
Accuracy
(% of reading)
-40 °C to 0 °C
(-40 °F to 32 °F)
±2.2 °C
0 °C to 260 °C
(32 °F to 500 °F)
±1.1 °C
November 2004
© 2004 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies
Question and Answer IconNeed help?

Do you have a question about the Fluke SureGrip 80PK-25 and is the answer not in the manual?

Fluke SureGrip 80PK-25 Specifications

General IconGeneral
MaterialStainless Steel
Temperature Range-40 to 260 °C
Sensor TypeThermocouple
HandleSureGrip
Resolution0.1 °C
Diameter3 mm
CompatibilityFluke Digital Multimeters and Temperature Probes

Summary

Introduction

Specifications

Type

Standard grade Ni-Cr vs Ni-Al (Chromel vs Alumel).

Measurement Range

Probe measurement range from -40 °C to 350 °C (-40 °F to 662 °F).

Accuracy

With respect to ANSI MC96.1 Standard Limits of Error, calculated in °C.

Probe Components and Ratings

Output and Seebeck Coefficient

Output voltage and Seebeck coefficient at specified temperatures.

Measurement Time

Time constant for step changes in measurement.

Maximum Voltage

Maximum voltage limits for probe operation.

Sheath Specifications

316 Stainless Steel, 3.4036 mm diameter, 10.16 cm length, conical tip.

Grounding

Junction welded to sheath for grounding.

Cable Specifications

40-inch PVC cable, gray jacket, max temp 105°C, food service limitation.

Conductor and Connector Details

Type K conductors, miniature connector with Hytrel handle.

Overall Probe Length

Total length of the probe from tip to cable strain relief.

Protection Class

Class 3, relating to insulation and grounding properties.

Measurement Considerations

Instrument Compatibility

Compatibility with Type K instruments, miniature connectors, and cold reference compensation.

Temperature Limitations

Maximum tip temperature and limitation of the handle end.

Exposure Limitations

Sheath material is stainless steel, avoid halides or sulfides.

Food Service Limitations

Conditions for food contact with probe and handle.

Operation Guide

Connecting, setting up, and checking the instrument.

Measuring Technique

Proper insertion depth to minimize shunting effect.

Troubleshooting

Steps for verifying instrument compatibility and probe continuity.

Scale Conversions

Celsius to Fahrenheit Conversion

Formula to convert Celsius to Fahrenheit.

Fahrenheit to Celsius Conversion

Formula to convert Fahrenheit to Celsius.

Contacting Fluke and Warranty

Contact Information

Phone numbers and website for application assistance and service.

Product Registration

Website for registering your Fluke product.

One Year Limited Warranty

Details of warranty coverage, exclusions, and remedy.

Related product manuals