Service Information
Performance Tests
7
Testing the Voltage Functions
The tests in this section check the following voltage functions:
Automatic selection of ac or dc voltage mode
Correct operation of the hazardous voltage indicator
Correct operation of the HOLD function
Perform the tests as follows:
1. Set the calibrator to the T5-600 or T5-1000 test voltage given in Table 2.
2. Apply the voltage to the tester. Verify that the tester reads within the display limits shown.
3. Apply the voltages and verify the testers readings for the steps in Tables 3 and 4.
Table 2. AC Voltage Test
T5-600 T5-1000
Calibrator Output
T5-600 Display Limits
(AC annunciator ON) Calibrator Output
T5-1000 Display Limits
(AC annunciator ON)
600 V, 60 Hz 589 to 611 1000 V, 60 Hz 983 to 1017
Table 3. DC Voltage Tests
T5-600 T5-1000
Step Calibrator Output
T5-600 Display Limits
(DC annunciator ON) Calibrator Output
T5-1000 Display Limits
(DC annunciator ON)
1 600 V dc 593 to 607 1000 V dc 989 to 1011
2 -60 V dc -58 to -62 -100 V dc -98 to -102
3 1.0 V dc 1 ±1* 1.0 V dc 1 ±1*
4 -1.0 V dc -1 ±1* -1.0 V dc -1 ±1*
* Verify that the ac annunciator is OFF.
Table 4. Hazardous Voltage LED and HOLD Function Tests
Step Calibrator Output T5-600/T5-1000 Display Indicators
1 12 V, 60 Hz Reading: 12 ± 1
Hazardous voltage LED OFF
2 12 V, 60 Hz
Tap testers HOLD button
Reading: 12 ± 1
Hazardous voltage LED OFF
HOLD indicator ON
3 +38 V dc Reading: 12 ±1
Hazardous voltage LED ON
HOLD indicator ON
w ww. . com
information@itm.com1.800.561.8187

Other manuals for Fluke T5-600

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke T5-600 and is the answer not in the manual?

Fluke T5-600 Specifications

General IconGeneral
WaterproofYes
Product typeDigital multimeter
AC current range0 - 100 A
AC voltage range0 - 600 V
Resistance range0 - 1000 Ω
Battery voltage1.5 V
Dimensions (WxDxH)30.5 x 51 x 203 mm
Weight and Dimensions IconWeight and Dimensions
Weight300 g

Related product manuals