General Information
Specifications
1
1-3
Response Time(Bargraph): Selectable; 24, 12, 6, and 3 seconds.
24 second response time selected:
Scale Range (CPM) Response (seconds)
1 0-100 24
2 0-1000 12
3 0-10K 3
4 0-100K 1.5
5 0-1M 1.5
6 0-10M 0.7
12 second response time selected:
Scale Range (CPM) Response (seconds)
1 0-100 12
2 0-1000 6
3 0-10K 3
4 0-100K 1.5
5 0-1M 1.5
6 0-10M 0.7
6 second response time selected:
Scale Range (CPM) Response (seconds)
1 0-100 6
2 0-1000 3
3 0-10K 1.5
4 0-100K 1.5
5 0-1M 0.7
6 0-10M 0.7
3 second response time selected:
Scale Range (CPM) Response (seconds)
1 0-100 3
2 0-1000 1.5
3 0-10K 0.7
4 0-100K 0.7
5 0-1M 0.7
6 0-10M 0.7