20
Our Control Solution contains a known amount of substance that reacts with
test strips and is used to ensure your device and test strips are working together
correctly.
Test strips, control solutions, or sterile lancets may not be included in the kit
(please check the contents on your product box). They can be purchased
separately.
Do a control solution test when:
9 you suspect the device or test strips are not working properly.
9 your test results are not consistent with how you feel, or if you think the results
are not accurate.
9 you have dropped or think you may have damaged the device.
To perform the control solution test, do the following:
1. Insert the test strip into the test slot of the device. Wait for the device to display
the test strip “ “ and blood drop “ “.
2. The meter will detect the difference between control solution and blood
samples automatically. It will automatically mark the result as a control solution
test with “QC” display.
3. Apply the control solution. Shake the control solution vial thoroughly before
use. Squeeze out a drop and wipe it off, then squeeze another drop and place
it on the tip of the vial cap. Hold the device to move the absorbent hole of the
test strip to touch the drop. Once the conrmation window is lled completely,
the device will begin counting down.