31
Total Cholesterol Test:
Message What it Means
< 100 mg/dL (2.65 mmol/L)
> 400 mg/dL (10.4 mmol/L)
Uric Acid Test:
Message What it Means
< 3 mg/dL (0.179 mmol/L)
> 20 mg/dL (1.190 mmol/L)
Error Message
Error Message
Cause What To Do
There is insufcient power
in the batteries to perform
the test.
Replace the batteries immediately and
reset the date and time on the meter
setting.
The strip has been used. Repeat the test with a new strip.
The blood sample has
been applied to the strip
before the meter was ready
for the measurement.
Repeat the test with a new strip.
Apply the blood sample when the
meter indicates that it is ready for the
measurement by displaying “ ”
and a
“ ” which starts blinking.