HOW TO PERFORM A CONTROL SOLUTION TEST
CAUTION :
Every time you perform a control solution test you must enter "QC" test mode so
that the test result will not be stored in the meter memory.
STEP 1. Insert test strip
Insert a test strip into the test slot with the contact bars end first and
facing up. (Contact bars must be inserted all the way into the meter or you
may get an inaccurate test result.) The meter turns on automatically and
displays the following in sequence:
When the "
" appears on the display, press the Main button and "QC" will
appear on the display. When the "QC" sign is displayed, the meter will not
store your test result in the memory. If you decide not to perform a control
solution test, press the Main button again and the "QC" sign will disappear.