Important Control Solution Information
Use only our control solutions with your meter.
Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution vial
and discard the remaining solution after 3 months.
It is recommended that the control solution test be done at room
temperature 20°C to 25°C (68°F to 77°F). Make sure your control
solution, meter,
and test strips are at this specified temperature range
before testing.
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Store the control solution tightly closed at temperatures between 2°C
to 30°C (36°F to 86°F). Do NOT freeze.