Important Control Solution Information
●
Use only our control solutions with your meter.
●
Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution vial
and discard the remaining solution after 3 months.
●
It is recommended that the control solution test be done at room
temperature 20°C to 25°C (68°F to 77°F). Make sure your control
solution, meter,
and test strips are at this specified temperature range
before testing.
●
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
●
Store the control solution tightly closed at temperatures between 2°C
to 30°C (36°F to 86°F). Do NOT freeze.