31
Step 4
Apply the sample.
Hold the blood drop to touch the absorbent hole of the test strip.
Blood will be drawn in and after the conrmation window is
completely lled, the meter begins counting down.
NOTE
Do not press the punctured site against the test strip or try to smear the
blood.
If you do not apply a blood sample to the test strip within 3 minutes, the
meter will automatically turn off. You must remove and reinsert the test
strip to start a new test.
The conrmation window should be lled with blood before the meter
begins to count down. NEVER try to add more blood to the test strip after
the drop of blood has moved away. Discard the used test strip and retest
with a new one.
If you have trouble lling the conrmation window, please contact your
health care professional or the local customer service for assistance.