Oxygen Concentrator User Manual
Oxygen Humidifier Bottle Checkings
If your doctor recommends using the device with humidifier;
• Check the humidifier bottle periodically, as sediment and lime will form in the water
tank during extended use.
• It is recommended to change the humidifier bottle once a month.
• Add drinking water (bottled water) or boiled water daily. Do not use tap water.
• Do not put more / less water than the indicated level.
• Make sure the lid is properly closed after adding water.
Cannula Checking
Make sure that the cannula used is not broken and not torsion. We recommend that
you change the cannula at the latest in 1 month depending on the intensity of use.
Filter Checking
Under the normal use conditions, it is recommended that the bacteria filter of device to be
changed once in a month. However, this time gets shorter using in very dusty environment.
The filter should be checked and changed frequently. This responsibility belongs to patient.
The device and the patient may be damaged by the blockage of the filter. If the device is
damaged due to failure to filter not being replaced on time, the device will be out of
warranty even it is under warranty term.
Changing the filter;
• Open the filter cover on the back of the device.
• Remove the filter and replace it. (Never try to clean the filter)
The sponge filter on the device (if any) should be checked frequently and removed
when it is contaminated. After washing with warm water, it should be dried and used again.
Worn filters should not be used.
The proper operation of the device is not only depends on device. Accessories are
very important. Checkings should be carried out as described above. Otherwise,
performance will decrease.
CALIBRATION
Operation of the device at the desired performance can be achieved by following
the instructions. In normal use, it is sufficient to calibrate the device once in 12 months. If it
is suspected that there is a decrease in the oxygen rate of the device, it should be verified
by measuring it with a calibrated oxygen analyzer.
CLEANING
Do not use petroleum based and solvent based products when cleaning the cabinet
of device. Clean the outer cabinet of device with alcohol through cloth. Thanks by this,
there will be no residue. Unplug the power cord from outlet while cleaning.
10 / 18