Oxygen Concentrator User Manual
Broadcast Test Compatibility Electromagnetic Environment- Guidance
RF Broadcast
CISPR 11
1. Group
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment
RF Broadcast
CISPR 11
Group B
The device is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage network that supplies buildings used for
domestic purposes.
Harmonic Emissions
61000-3-2
Group A
Voltage Change/Vibrating
Emissions
61000-3-3
Compliance
COMPLIANCE WITH STANDARTS
IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, ISO
80601-2-67, ISO 80601-2-69 standarts are referenced in the production of oxygen
concentrators.
IEC 60601-1 CLASSIFICATION
Protection type against electric shock : Class II
Protection degree against electic shock : Type BF
Degree of protection against impact and waterproof : IP 21
SOFTWARE INFORMATION
Firmware is included in class A software security class. Class A: Does not cause
injury or damage to health. The version definition is used for traceability of software
updates. Updates are performed on all models. The current version of oxygen
concentrators is 1.0.
Software Difference: OXY300 model oxygen concentrator has the same visuality, same
hardware and the same physical characteristics as the OXY500 model as it goes through
the same manufacturing processes. The difference between them is software and the
OXY300 model has the capacity to give 3 liters of oxygen per minute, while the OXY500
model has the ability to give 5 liters of oxygen per minute.
Software Operation: : In both models, when the power cable of the device is plugged in
and the turn on button is pressed, the device starts in 1 second. Then it starts operating
and continues to supply oxygen to the patient at the adjusted flow.
17 / 18