Oxygen Concentrator User Manual
ALARMS
Low : When that audible and visual alarm emerges on device, it indicates that
the oxygen level decreases below %82..
Call Service : The device gives an audible and visual alarm in the event of a power
failure and when the flowmeter is 0 (zero).
TECHNICAL SPECIFICATIONS
OXY300 OXY500
Oxygen Level 1-3 LPM Oxygen Level 1-5 LPM
Pressure 8,5 psi (58.6 kpa) Pressure 8,5 psi (58.6 kpa)
Watt 300W Watt 300W
Electrical Spec. 230V, 50Hz, 1.8A(max) Electrical Spec. 230V, 50Hz, 1.8A(max)
Oxygen
Percentage
1~3 LPM 94%±3% Oxygen
Percentage
1~5 LPM 94%±3%
Fuse 15A Fuse 15A
Av. O² Purity 1 lt/min. : %94±3%
2 lt/min. : %94±3%
3 lt/min. : %94±3%
Av. O² Purity 2 lt/min. : %94±3%
4 lt/min. : %93±3%
5 lt/min. : %92±3%
Operating
Conditions
Temp. :10ºC / 40ºC
Humidity : %30 / %70
Pressure : 50/106kpa
Operating
Conditions
Temp. :10ºC / 40ºC
Humidity : %30 / %70
Pressure : 50/106kpa
Weight 14kg Weight 15kg
Sound Level <48dbA Sound Level <48dbA
Dimensions 400*300*510mm Dimensions 400*300*510mm
Storage
Conditions
Temp. :-20ºC/ 50ºC
Humidity : %0 / %95
Pressure : 50/106kpa
Storage
Conditions
Temp. :-20ºC / 50ºC
Humidity : %0 / %95
Pressure : 50/106kpa
O² Capacity 3 Liter O² Capacity 5 Liter
Model Diff. Continuously 3 lt Model Diff. Continuously 5 lt
Working System Producing electric withO² Working System Producing electric withO²
Class/Protection
type
II / BF Class/Protection
type
II / BF
Expected
Operating Life
10 years Expected
Operating Life
10 years
Warranty 2 years Warranty 2 years
IP Value IP21 IP Value IP21
Alarm Available Alarm Available