Oxygen Concentrator User Manual
ENVIRONMENTAL CONDITIONS
• Do not use oxygen concentrator devices in any toxic environment.
• Do not use oxygen concentrator devices in environmental conditions where
explosive and chemical gases or other flammable anaesthetic agents are present.
Keep fire away.
• Do not use under rain, direct sunlight, in extremely humid-dusty environment or in
smoking areas.
• Storage Conditions: Temperature -20ºC / 50 ºC, Humidity %0 - %95, Pressure
50kPa/106kPa
• Operating Conditions: Temperature 10ºC / 40 ºC, Humidity %30 - %70, Pressure
50kPa/106kPa
• The device performance will decrease in case of storing out of abovementioned
temperature, humidity and pressure limits.
TROUBLESHOOTING TABLES
Trouble Solution
Device does not deliver
oxygen
* Make sure that the power cord of device is plugged to
outlet.
* Make sure that device is operating.
* Make sure the flowmeter is set as required.
* Check the connections of nasal cannula.
* Make sure the lid of humidifier bottle is closed properly.
* Make sure the tube inside humidifier bottle is not
blocked.
* Contact technical service
Device gets overheat
* Make sure that device is not close to wall.
* Make sure the filters are not dirty.
* Pay attention that the environment is not too hot where
the device is used.
* Pay attention to the ventilation of the environment where
the device is used.
* Contact technical service.
Device emerges O² Low
alarm
* Ventilate the environment where device is located.
* Contact technical service.
Device shows Call Service
alarm
* Make sure that electrical power is supplied to the device.
* Make sure flowmeter setting is not at 0 (zero) or over 5.
* Contact technical service.
Device operates noisily * Operate the device on flat surface.
* Contact technical service.
No air flow on device
* Put your finger to the oxygen outlet of the device.
* Check the flow showed on display.
* Contact technical service.
12 / 18