Oxygen Concentrator User Manual
ENVIRONMENTAL CONDITIONS
Do not use oxygen concentrator devices in any toxic environment.
Do not use oxygen concentrator devices in environmental conditions where
explosive and chemical gases or other flammable anaesthetic agents are present.
Keep fire away.
Do not use under rain, direct sunlight, in extremely humid-dusty environment or in
smoking areas.
Storage Conditions: Temperature -20ºC / 50 ºC, Humidity %0 - %95, Pressure
50kPa/106kPa
Operating Conditions: Temperature 10ºC / 40 ºC, Humidity %30 - %70, Pressure
50kPa/106kPa
The device performance will decrease in case of storing out of abovementioned
temperature, humidity and pressure limits.
TROUBLESHOOTING TABLES
Trouble Solution
Device does not deliver
oxygen
* Make sure that the power cord of device is plugged to
outlet.
* Make sure that device is operating.
* Make sure the flowmeter is set as required.
* Check the connections of nasal cannula.
* Make sure the lid of humidifier bottle is closed properly.
* Make sure the tube inside humidifier bottle is not
blocked.
* Contact technical service
Device gets overheat
* Make sure that device is not close to wall.
* Make sure the filters are not dirty.
* Pay attention that the environment is not too hot where
the device is used.
* Pay attention to the ventilation of the environment where
the device is used.
* Contact technical service.
Device emerges O² Low
alarm
* Ventilate the environment where device is located.
* Contact technical service.
Device shows Call Service
alarm
* Make sure that electrical power is supplied to the device.
* Make sure flowmeter setting is not at 0 (zero) or over 5.
* Contact technical service.
Device operates noisily * Operate the device on flat surface.
* Contact technical service.
No air flow on device
* Put your finger to the oxygen outlet of the device.
* Check the flow showed on display.
* Contact technical service.
12 / 18