999075 Rev F Page 8 of 40
A risk assessment must be performed before using any other manufactured sling, carry bar or ceiling
track to ensure ‘safe’ use can be established.
Risk of strangulation: Please make sure handset cable and lift tape are always clear of all persons.
Risk of impact with carry bar: Please take care to ensure the carry bar is clear of the person in the sling
when preparing to raise/lower and move them to avoid any contact with that person.
Risk of collision: The person operating the hoist should make sure that when raising, lowering, or moving
the hoist that no people or objects will obstruct, be injured or damaged by the movement.
Ensure that the person being hoisted is always raised clear of the floor when using the hoist.
Ensure the lift tape is vertically aligned with the hoist when raising or lowering the carry bar. Any
deviation from this can cause the tape to fray and result in its potential failure.
Serious Injury: If, during the use of this device or because of its use a serious incident has occurred,
please report it to the manufacturer and to your national authority.
Electric Shock: Do not insert any objects into the hoist case or battery charging station because of
potential risk of electric shock. To reduce the risk of electric shock, do not install or operate the battery
charger with a damaged cable or if the unit has been dropped or damaged.
Portable RF Communication Devices: Portable RF communications equipment (including peripherals,
such as antenna cables and external antenna) should be used no closer than 30cm (12 inches) to any
part of the Ceiling Track Hoist, including cables specified by the manufacturer, otherwise degradation of
the performance of this equipment could result.
Vicinity to Other Equipment: Use of this equipment adjacent to or stacked with other equipment should
be avoided, as it could result in improper operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are operating normally.
Specified Accessories: Use of accessories, transducers, and cables other than those specified or
provided by the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
You may need to seek specialist advice on how to assist some people with specific moving and
handling needs. Sources of advice include, but is not limited to, professional bodies and organisations,
occupational therapist, physiotherapists, manual handling advisers and ergonomist with experience in