INF-TN2FSC-E 177

Flow velocity prole can be observed in real time using the dedicated detector from the outside. It is speciable by the
code symbol of ow transmitter.
APPLICATION
Pulse Doppler method is applicable to observe ow velocity prole in real time, display the ow status in the pipe, and
decide the appropriate measurement location. Also, it can be used for diagnosis of ow and laboratory test.
SPECIFICATIONS
Measuring uid: Uniform liquid in which ultrasonic waves can propagate.
Turbidity of uid: Axisymmetric ow in a lled pipe.
Fluid temperature: 40 to +100°C (FSDP2)
−40 to +80°C (FSDP1,FSDP1)
Air bubble quantity: 0.02 to 15Vol% (Velocity is 1m/s)
Pipe size: Small type sensor : φ40 to φ200mm
Middle type sensor :φ100 to φ400mm
Large type sensor :φ200 to φ1000mm
Measurement range: 0 to ±0.3: ±Maximum Velocity (depending on the pipe diameter) Refer to chart, page 145.
Note) This function is to observe ow velocity prole, and it may be different from actual
ow rate.

Mounting method: Mounting on outside of existing pipe
Ambient temperature: −20 to +80°C
Ambient humidity: 100% RH or less
Type of enclosure: IP67 (with waterproof BNC connector provided.)
Material: Sensor housing: PBT
Guide frame: Aluminum alloy
Mounting belt: Plastic cloth belt/stainless wire