INF-TN2FSC-E 177
Flow velocity prole can be observed in real time using the dedicated detector from the outside. It is speciable by the
code symbol of ow transmitter.
APPLICATION
Pulse Doppler method is applicable to observe ow velocity prole in real time, display the ow status in the pipe, and
decide the appropriate measurement location. Also, it can be used for diagnosis of ow and laboratory test.
SPECIFICATIONS
Measuring uid: Uniform liquid in which ultrasonic waves can propagate.
Turbidity of uid: Axisymmetric ow in a lled pipe.
Fluid temperature: −40 to +100°C (FSDP2)
−40 to +80°C (FSDP1,FSDP1)
Air bubble quantity: 0.02 to 15Vol% (Velocity is 1m/s)
Pipe size: Small type sensor : φ40 to φ200mm
Middle type sensor :φ100 to φ400mm
Large type sensor :φ200 to φ1000mm
Measurement range: 0 to ±0.3: ±Maximum Velocity (depending on the pipe diameter) Refer to chart, page 145.
Note) This function is to observe ow velocity prole, and it may be different from actual
ow rate.
Mounting method: Mounting on outside of existing pipe
Ambient temperature: −20 to +80°C
Ambient humidity: 100% RH or less
Type of enclosure: IP67 (with waterproof BNC connector provided.)
Material: Sensor housing: PBT
Guide frame: Aluminum alloy
Mounting belt: Plastic cloth belt/stainless wire