3 - 7
3.4 Sampling
3.4.1 Conditions of sampling gas
'XVWFRQWDLQHGLQWKHVDPSOLQJJDVVKRXOGEHFRPSOHWHO\UHPRYHGZLWKD¿OWHU)RUWKH¿QDO
VWDJH¿OWHUXVHD¿OWHUWKDWDOORZVUHPRYLQJGXVWSDUWLFOHVRIP
'HZSRLQWRIVDPSOLQJJDVPXVWEHORZHUWKDQWKHDPELHQWWHPSHUDWXUHWRDYRLGRFFXUUHQFH
of drain in the gas analyzer. If vapor is contained in the sampling gas, dew point should be
ORZHUHGWRÛ&E\XVLQJDGHKXPLGL¿HU
,I62
PLVWLVFRQWDLQHGLQWKHVDPSOLQJJDVXVHDPLVW¿OWHURUFRROHUWRUHPRYH62
mist.
2WKHUPLVWVVKRXOGEHUHPRYHGE\XVLQJDPLVW¿OWHURUFRROHU
&RUURVLYHJDVHVVXFKDV&O
2
, F
2
DQG+&OLIWKH\DUHFRQWDLQHGLQWKHVDPSOLQJJDVLQ consider-
able amounts, will shorten the life of instruments.
7HPSHUDWXUHRIVDPSOLQJJDVVKRXOGEHZLWKLQWRÛ&3URYLGHDPHDQVWKDWSUHYHQWVHQWU\
of hot gas directly into the instrument.
6DPSOLQJJDVÀRZ
)ORZRIVDPSOLQJJDVVKRXOGEH/PLQ/PLQ
$YRLGÀRZÀXFWXDWLRQGXULQJPHDVXUHPHQW
2EVHUYHWKHÀRZUHDGLQJE\DÀRZPHWHUSURYLGHGDVVKRZQLQWKHH[DPSOHRIWKHVDPSOLQJV\V-
WHPFRQ¿JXUDWLRQ6HFWLRQ
3.4.3 Preparation of standard gas
5RXWLQHFDOLEUDWLRQLVUHTXLUHGE\VWDQGDUGJDVIRUNHHSLQJWKLVLQVWUXPHQWXQGHUQRUPDORSHUDWLRQ
condition (once a week). Prepare a standard gas cylinder for zero calibration and span calibration.
Analyzer without O
2
measurement
Analyzer with built-in O
2
sensor
Analyzer with external
zirconia O
2
sensor
=HURJDV N
2
gas or dry air N
2
gas 'U\DLUDWPRVSKHULFDLURU
gas with a concentration of
RUPRUHRIIXOOVFDOH
Span gas other than
for O
2
measurement
Gas with concentration of
RUPRUHRIIXOOVFDOH
Gas with concentration of
RUPRUHRIIXOOVFDOH
Gas with concentration of
RUPRUHRIIXOOVFDOH
Span gas for O
2
measurement
Gas with concentration of
RUPRUHRIIXOOVFDOH
RUDWPRVSKHULFDLU
WR2
2
,QWKHFDVHRI&2
2
DQDO\]HU]HURJDVVKDOOQRWFRQWDLQ&2
2
JDVLIWKHFRQFHQWUDWLRQRI&2
2
in at-
mospheric air is considerable.
3.4.4 Purging of instrument inside
The inside of instrument need not be purged generally except for the following cases.
(1) A combustible gas component is contained in sampling gas.
&RUURVLYHJDVLVFRQWDLQHGLQWKHDWPRVSKHULFDLUDWWKHLQVWDOODWLRQVLWH
7KHVDPHJDVDVWKHVDPSOHJDVFRPSRQHQWLVFRQWDLQHGLQWKHDWPRVSKHULFDLUDWWKHLQVWDOOD-
tion site.
In such cases as above, the inside of analyzer should be purged with the air for instrumentation
or N
2
.
3XUJLQJÀRZUDWHVKRXOGEHDERXW/PLQ
If dust or mist is contained in purging gas, it should be eliminated completely in advance.