What causes baseline lead record on Fukuda Denshi Medical Equipment?
- EEdward PhillipsSep 12, 2025
If a certain lead record appears as a baseline on your Fukuda Denshi Medical Equipment, check the patient cable for any breaks and verify its connection status. Also, inspect the operation of the OP Amp, IC23 to IC27, and A/D converter IC37 (related to the baseline lead) and the transistors Q13 to Q21 for protection against excess input to ensure they are not shorted.