MODEL 610
Liquid Nitrogen
Battery Operated Low Level Alarm
16380 13 Mile Rd. • Roseville, Michigan 48066 • (586) 778-0426 • Fax (586) 778-0479
www.gordinier.com Email: pat@gordinier.com
Question and Answer IconNeed help?

Do you have a question about the GE 610 and is the answer not in the manual?

GE 610 Specifications

General IconGeneral
BrandGE
Model610
CategorySecurity System
LanguageEnglish

Summary

Product Receiving and Unpacking

Inspect Shipping Carton and Contents

Examine shipping carton for damage. Contact distributor for claims and check for apparent shipping damage like broken knobs or dented cases.

Model 610 Key Features

Product Overview and Features List

Economical, battery-operated alarm with built-in charger, test circuit, and probe protection, offering up to 5 days of alarm.

Mounting the Controller Unit

Attach Bracket and Controller

Secure the mounting bracket to the storage unit handle using the existing screw and attach the controller with Velcro film.

Sensor Probe Installation Guide

Position Probe in Liquid Nitrogen Tank

Lower sensor probe to tank bottom, raise to desired level (2-3 inches suggested) for safety margin. Consult manufacturer for holding time data.

Connecting Probe and Power

Connect System Components

Plug the low sensor probe and the 9 Volt AC transformer connector into the controller. Ensure proper grounding for AC outlet.

Optional Special Order Lid

Special Order Lid Diagram

Diagram illustrating the special order lid and its insertion into the dewar, with a reference to probe connection details.

Controller Testing and Battery Charging

Test Alarm Functionality

Test alarm by removing sensor from liquid nitrogen or using the "Push to Test" button. Note audible signal behavior.

Battery Charging Procedure

Recharge batteries monthly (or weekly) using the 9V AC adapter. Recharging takes about 15 hours; a light indicates charging status.

Low Level Alarm Calibration Procedure

Perform Calibration Steps

Submerge probe in LN2, use toggle switch and adjuster to set tone disappearance, then test calibration.

Unit Care and Maintenance Guidelines

General Care Recommendations

Keep unit dry, handle gently, use in normal temperatures, avoid dust, and clean with a damp cloth.