CALIBRATING LOW LEVEL ALARM
MODEL 610
Model  610                                                                                                                                        8
•  Submerge the Probe in liquid nitrogen (LN2) for 10 minutes.
•  Press the battery “On/Off-Cal” toggle switch downward and hold.
•  Use  the screwdriver (supplied) turn the adjuster clockwise or counter-clockwise
    until the tone  begins to sound.  Slowly turn the adjuster until the tone sound just disappears.
•  Release the “Cal” toggle switch to its battery “Off” position.
•  Press “Cal” toggle switch to its full upright position or battery “On”.
•  Press the “Push to Test” button to assure the battery power is turned on and calibrated.
 ADJUST WITH
SCREW-
DRIVER
“ON/OFF”-CAL
TOGGLE SWITCH
 PROBE
 TEST BUTTON