ÄKTA start Maintenance Cue Card, 29-0240-43 AC
5
Cleaning system ow path
Purpose
The system methods are used for cleaning the system ow
path, including all the inlets and outlets. The following wash
methods are available:
Pump wash A, Pump wash B
Washout fractionation tubing
System cleaning
Pump wash A/B
The Pump wash method is used to wash the selected
inlets/outlets and ow path tubing at high ow rate using
the specied buer port. Pump wash prevents cross-
contamination and carryover of buers/samples.
The Pump wash method is recommended:
Before starting a run when the system is not in use for a
prolonged period of time
When changing buers
Instruction
Immerse the buer A or B inlet tubing in DM water or
buer of choice.
Follow the workow below to run Pump wash.
Operation from the Instrument Display
Perform the following operations to run Pump wash.
•
•
•
•
•
1.
2.
Select
method
Method run
Prepare system
Pump wash A
End
Run
Exit
Pump wash B
Immerse
inlet A
in Buffer A
Immerse
inlet B
in Buffer B
ÄKTA start
Interval Maintenance action
When
required
Clean the Fraction collector.
Note: Frac30 is not a spill-free design. There
will be some spillage when tubes shift
position, depending on the ow rate.
Perform System cleaning, Pump Wash A
and B, Washout Fractionation tubing.
Clean the UV ow cell.
Clean the Conductivity cell.
Calibrate UV.
Calibrate Conductivity.
Perform Zero oset for the Pressure sensor.
Replace the inlet lters.
Replace the tubing and connectors.
Yearly,
or when
required
Check the drive sleeve on the Fraction
collector. Replace if worn out.
Operation from UNICORN start
Perform the following operations to run Pump wash.
UNICORN start
Method run
System control
Pre defined
System preparation
Pump wash A /
Pump wash B
Washout fractionation tubing
The Washout fractionation tubing method is used
to clean the fractionation tubing. The method is
recommended:
Before collecting the fractions using the Outlet valve
without Fraction collector
When changing buers
•
•
Instruction
Immerse the buer inlet tubing in DM water or the
buer of choice.
Disconnect the column from the ow path and
reconnect the tubing with a Union connector.
Follow the workow below to run Washout
fractionation tubing.
1.
2.