GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-il
X-RAY
UNIT
8-3179
SM
A0654E
SECTION
12.
PRELIMINARY
KVP,
TIMER,
AND
LOAD
CHECKS
Initial
Conditions:
a.
Circuit
breaker
and
keyed
switch
OFF.
=
b.
Tube
unit
shielded
and
x-ray
port
blocked
(from
Section
10).
НН
|
Hi
i
c.
H.
V.
Divider
connected
(from
Section
10).
il
ply
1444
Instrumentation:
|
+
a.
Connect
scope
to H.
V.
Divider,
refer
to
I
Mustration
4-11-30.
A
Scope
to
be
grounded
through
line
cord
and
scope
ground
connected
to
AMX-110
25282-P10
chassis.
Use
external
trigger
circuit
per
,
iNustration
4-11-31.
Or
INT,
DC,
+.
ILLUSTRATION
411-33
b.
Digital
voltmeter
across
battery
voltage
Timer:
80
KVP,
10
MAS
(from
Section
11).
У
=
20
KV/DIV
H
=
10
MS/DIV
_
Xray
tub
t
itored
by
0-150
DCMA
TRIG
=
Ext.
Trigger
Circuit
or
Int,
DC,
T
é
meter
(from
‘Section
11
or
Y
NOTE:
Deep
discharge
pulses
as
shown
above
appear
on
scopes
with
faster
writing
rates.
12-1
Timer
Check
And
Adjustment
Older
scopes
wil!
appear
as
in
Ill.
4-11-34.
1.
Turn
circuit
breaker
and
keyed
switch
ON.
+
4,82.
Select
80
KVP,
10
MAS.
Make
an
exposure
and
+
record
exposure
duration
from
scope
waveform
-
+
measure
time
interval
at
75%
of
maximum
KVP
(60
KVP
for
a
selected
80
KVP).
See
Ill.
4-11-33.
teto
ed
eo
4,43.
Select
80
KVP,
100
MAS.
Make
another
ex-
posure
and
note
exposure
duration
on
scope.
|
3
See
(If.
4-11-34.
3
4.
Ifnecessary
adjust
3A1R7
for
exposure
interval
| |
?
| |
|
within
+/-2%
of
.10
second
at
10
mAs
and
21514P1Q
1.0
seconds
at
100
mAs.
a
ILLUSTRATION
4-11-34
5.
Select
80
kVp,
1
mAs.
Make
another
exposure
.
and
only
if
necessary,
adjust
3A1R9
for
Timer:
80
KVP,
100
MAS
exposure
interval
of
10
MS.
V
=
20KV/DIV
H
=
200
MS/DIV
A6.
Turn
keyed
switch
and
circuit
breaker
OFF.
TRIG
=
Ext.
Trigger
Circuit
411-29