GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-ll
X-RAY
UNIT
10-377
12-2
Back-Up
Timer
Check
1.
Temporarily
disconnect
the
MAS
selector
lead
at
3TB1-21
and
tape
the
lead.
2.
Turn
circuit
breaker
and
keyed
switch
ON.
2.83,
Select
50
KVP,
and
MAS,
and
make
an
exposure.
Measure
exposure
duration
with
watch
or
on
scope
KVP
waveform
with
horizontal
time
of
0.5
sec/div.
4.
Exposure
should
be
terminated
within
3-1/2
seconds
by
back-up
time
delay
relay,
3K3.
5.
Turn
keyed
switch
and
circuit
breaker
OFF.
Reconnect
MAS
switch
lead
to
3TB1-21.
6.
Record
Data
а.
F3382
HEW
Test
No.
25.
b.
3316
Functional
Check
No.
5
and
13.
12-3
Battery
Load
Voltage
Check
1.
Turn
circuit
breaker
and
keyed
switch
ON.
sa
2.
Make
an
exposure
at
80
KVP,
200
MAS
and
note
battery
voltage
drop
between
standby,
stator
load,
and
x-ray
load.
3.
Voltage
drop
from
standby
to
stator
load
should
be
from
1
to
1.5
volts
and
from
standby
to
x-ray
load
up
to
20.6
to
24
volts.
If
voltage
drop
is
in
excess
of
25
volts
check
tightness
of
primary
terminal
and
connections
to
battery.
|f
necessary
check
each
cell
group
voltage
under
x-ray
load
to
locate
and
correct
cause
of
excessive
voltage
drop.
“a
4.
Make
a
series
of
exposures
at
2
MAS,
from
80
to
110
KVP.
Observe
scope
waveform
for
signs
of
arcing
or
transients.
5.
Turn
keyed
switch
and
circuit
breaker
OFF.
Do
not
disconnect
any
test
equipment.
411-30
SM
AO654E
21815-P10
ILLUSTRATION
4-11-35
Back-up
Timer:
50
КУР
V
=
20KVP/DIV
H
=
500
MS/DIV
TRIG
=
Ext.
Trigger
Circuit
Ra
21816-P10
ILLUSTRATION
4-11-36
80/90/100/110
KVP;
2
MAS
V
=
20KVP/DIV
H
=
2MS/DIV
TRIG
=
Ext.
Trigger
Circuit

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX-II and is the answer not in the manual?

GE AMX-II Specifications

General IconGeneral
BrandGE
ModelAMX-II
CategoryMedical Equipment
LanguageEnglish

Related product manuals