GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-II
X-RAY
UNIT
10-377
SM
A0654E
e.
Retighten
the
two
light
housing
mounting
screws
and
replace
the
namepiate
and
con-
trol
knobs.
+.
|
Repeat
the
alignment
check
from
Step
1.
10.
If
the
indicated
versus
actual
field
size
does
not
fail
within
specifications,
adjust
the
appropriate
pointer
or
pointers
as
follows:
a.
Remove
the
Blade
Control
Knobs
and
the
nameplate.
Refer
to
Section
14-1.
b.
Gently
bend
the
pointer
sideways
to
make
the
necessary
adjustment.
c.
Replace
the
nameplate
and
the
knobs.
d.
Repeat
the
alignment
check
from
Step
1.
LIGHT
11.
Record
collimator
Beam
Alignment
on
F-3382,
CREWS
Sec.
7
Test
SET-F.
o
=?
m
3
22009-P10
411-39