M1110147 Rev. 5 © 2007 Datex-Ohmeda. All rights reserved.  6-3 
 
Symptom  Likely causes  Troubleshooting steps 
Hour meter is not 
working. 
Fiberoptic cable is 
not fully inserted 
into the light box. 
Hour meter 
failure. 
1.  The hour meter only runs when the 
fiberoptic light pad is illuminated.  If the 
fiberoptic cable is not fully inserted in the 
light box, the LEDs are automatically shut 
off and the hour meter does not run.  
2.  If the fiberoptic cable is fully inserted and 
the hour meter still does not work, 
remove the unit from service and seek 
assistance from service personnel. 
No light output.  Fiberoptic cable is 
not fully inserted 
in the light box. 
 
Unit is not 
powered. 
1.  Confirm that the fiberoptic cable is fully 
inserted in the light box. If the fiberoptic 
cable is not fully inserted in the light box, 
the LEDs are automatically shut off. 
2.  If the front panel standby switch is in the 
“on” position and the small green LED on 
the switch is lit and there is still no light to 
the fiberoptic light pad, remove the unit 
from service and seek assistance from 
trained service personnel. 
3.  If the front panel standby switch is in the 
“on” position and the small green LED on 
the switch is not lit, then confirm that the 
unit is plugged into a powered outlet. 
Fiberoptic light pad 
connection is too 
loose. 
“Breakaway” 
feature. 
1.  The fiberoptic cable connection is 
designed with a “breakaway” feature to 
protect the unit if the fiberoptic cable is 
pulled.  If the fiberoptic cable has slack 
and the fiberoptic cable connector will not 
stay connected, remove the unit from 
service and seek assistance from trained 
service personnel.