9-2                          M1110147 Rev. 5 © 2007 Datex-Ohmeda. All rights reserved. 
 
Performance Specifications 
Spectral Irradiance (bare fiberoptic pad)* 
 
 
 
 
*Using an Ohmeda Medical BiliBlanket 
Meter II 
Large Fiberoptic Pad – 49 µW•cm
-2
•nm
-1 
(+/- 
25%) 9-point check  
 
Small Fiberoptic Pad – 70 µW•cm
-2
•nm
-1 
(+/- 
25%) 6-point check 
Note: When the BiliSoft fiberoptic light pad is 
inserted into a BiliSoft Pad Cover or BiliSoft 
Nest, the nominal spectral irradiance is 35 
µW•cm
-2
•nm
-1
 (large pad) and 50 µW•cm
-
2
•nm
-1
 (small pad) 
Wavelength  430-490 nm (peak 440-460 nm)  
LED module estimated life
*
Under continuous use, tested at room 
temperature, a typical LED module will run for 
approximately 8,000 to 10,000 hours before 
the light intensity drops 25%. 
Sound level  < 44 dB(A) at 1 meter 
X-ray X-ray compatible 
Physical Specifications 
Light box (W x H x L)  16.5 x 21 x 16.5 cm 
Light box weight (excluding fiberoptic pad)  < 2.5 kg 
Fiberoptic Pad weight  < 1.1 kg 
Fiberoptic light pad, small  15 x 30 cm (light-emitting area) 
Fiberoptic light pad, large  25 x 30 cm (light-emitting area) 
Fiberoptic cable length  137 ± 5 cm 
Regulatory Standards 
IEC Type B equipment.  
IEC Class 1 (continuous operation) 
FDA Class II 
Product certified to the following 
standards: 
EN60601-1; EN60601-1-2; EN60601-2-50; ISO 
10993-5; ISO 10993-10; UL 60601-1; CSA 
C22.2 No 601.1-M90; IEC 60601-1-8; BS EN 
980; 16CFR Part 1632.6 (for BiliSoft Pad 
Covers and BiliSoft Nests) 
*
The LED module life may vary when used in the actual clinical environment.  Factors such as duty cycle and 
ambient temperature may impact the life of the LED module.  Measure the irradiance of the BiliSoft System and 
replace the LED module when the system is below specifications.