5
Turning the CARESCAPE ONE On/O
Turning On the CARESCAPE ONE
CS ONE can operate via battery power using its own
battery, or via DC power when connected to a powered
docking station.
1. Ensure all cables are properly connected.
2. Press the on/standby button on the handle side of
the CS ONE. The welcome screen will appear.
Turning O the CARESCAPE ONE
1. Ensure that the patient has been discharged/the
case has ended.
2. Press the on/standby button for approximately
1 second.
The Powering Down window appears to conrm
you want to shut down the CS ONE.
3. Press the on/standby button again within 10
seconds to shut down the CS ONE.
If you do not press the on/standby button again
within 10 seconds, the CS ONE remains on and the
Powering Down window is removed from
the display.
Inserting and Removing
a Battery
Caution: To prevent loss of monitoring, or the
CS ONE from powering down, change the CS ONE
battery when the CS ONE is connected to a powered
CS ONE dock F0.
Inserting a Battery
1. Open the battery door by gently pulling on the
battery door pull tab.
2. Position the battery with the connector end facing
towards the battery slot and insert the battery all
the way into the battery slot.
Removing a Battery
1. Open the battery door by gently pulling on the
battery door pull tab.
2. Pull on the battery cord to remove the battery from
the battery slot.
2
1
2