28
Additional alarm settings
The following alarms can be turned o to reduce
nuisance alarms during manual ventilation if clinically
appropriate:
1. CO
2
Alarms: Use the CO
2
Alarms setting to turn o
the CO
2
‘Apnea’ alarm, EtCO
2
low, EtCO
2
high, and
FiCO
2
high and Absorbent OK? alarms during
manual ventilation.
2. Vol Apnea Alarm: Use the Vol Apnea Alarm setting
to turn o the volume apnea alarm during manual
ventilation.
3. MV/TV Alarms: Use the MV/TV Alarms setting to
turn o the MV and TV alarms. If MV/TV Alarms is
set to O during manual ventilation, the alarms
remain o until the Bag/Vent switch is set to Vent or
the MV/TV Alarms is set to On. If the MV/TV
Alarms is set to O during mechanical ventilation,
the alarms remain o when manual ventilation
starts.
Note: CO
2
Alarms and Apnea Alarm selection will not
be visible unless enabled by a Super User. CO
2
Alarms
and Vol Apnea Alarm can only be turned o while in
manual ventilation.
1 2 3