36
ecoFLOW menu components
1. Agent: Shows Agent cost and ow information.
2. Agent Cost: The cost of the current agent ow.
This value is determined by the agent ow multiplied
by the agent cost set by the Super User.
3. FiO
2
owmarker: The graphical representation on
the ow tube of the FiO
2
ow value. This marker can
be removed by the Super User
4. Agentow: The measured value of the liquid agent
ow from the vaporizer. The agent ow may have a
delayed response.
5. FiO2ow:The minimum O
2
ow needed to
maintain the set inspired O
2
ow. This item can be
disabled by the Super User.
6. O
2
total (numeric): The O
2
total equals the set of
O
2
ow. If N
2
O is the balance gas, this equals the set
O
2
ow. If Air is the balance gas, this is the set of O
2
ow plus 21% of the Air ow.
7. O
2
total (graphical): The graphical representation
of the total O
2
ow.
8. N2O total and N2 total: The N
2
O total equals the
set N
2
O ow. The N
2
total equals the N
2
in the Air
ow.
9. Flow bobbin: The height of this represents the total
fresh gas ow delivered to the breathing system.
10.Totalow: Shows Total Flow information.
I/min ml/h
02 Total
Sev
€ 6.00 /h
Fi 35
Fi 35 02 flow
0.50
1
2
200
9
20
15
3
Total Flow
0.55
8
Agent
5
10
6
7
8
9
1
2
3
4