Weight
Unpacked, without input and output
table
45,0 kg
Unpacked, with input and output
table 43 cm
49,4 kg
Unpacked, with input and output
table 43 cm, incl. light cover
50,5 kg
Pixel size
High resolution 35 µm
Medium resolution 75 µm
Electrical Connection
Operating voltage Auto ranging power supply from: 90 V to 264 V,
accuracy + 10%
Class I with protection earth
Connect to earthed supply circuitry only.
Mains frequency 50/60 Hz
Current rating Max. 2 A
Mains fuse protection Europe: min. 10 A, max. 16 A
USA & Japan: min. 10 A, max. 15 A
Digitizer input voltage / current 24 V DC / max. 6,25 A
Network connectivity
Ethernet connector RJ45 female, 10/100 Mbit/s autosensing,
shielding CAT5
Power consumption
Standby 110 V - 240 V / 50-60 Hz
configuration
max. 22 W
During operation 110 V - 240 V / 50-
60 Hz configuration
max. 140 W
Environmental conditions CRxVision during operation
Room temperature recommended: 20 °C - 25 °C
allowed: 15 °C - 30 °C
Maximum temperature change < 0.2 °C/min.
Relative humidity recommended: 30 % - 60 %
allowed: 10 % - 85 %, non condensed
Light conditions Must be operated in a dark room.
CRxVision Operating Manual V1.2 43