EasyManua.ls Logo

GE Druck DPI 740

GE Druck DPI 740
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
K0200 Issue No. 1
2 D PI 7 4 0 U se r M a n u a l
1.1 Specification
Dimensions
Weight: (nominal).................................................................. .................. 0.5 kg
Size: height = 190 mm width = 90 mm, depth = 36 mm
Environmental
Temperature:
Operating .......................... ........ ......... ......... -10 to +50
C
Storage ......................................................... ......... -40 to +70
C
Pressure medium:................ Any gas compatible with pyrex,
.................. .................. silicon, stainless steel and epoxy resin
Sealing standard................................................................... .................. IP54
Pressure
Connector...................................................... ......... 6 mm o/d
or.................................................... ......... 4 mm i/d hose fitting
Accuracy
Non-linearity, hysteresis and repeatability:
over the range of 10° to 30°C
........................................................................... ......... ±0.02 % FS
over the range of 0° to 40°C
........................................................................... ......... ±0.03 % FS
over the range of -10° to 50°C
........................................................................... .........±0.045 % FS
Stability .................................................................. ........ <100 ppm/year
Range (barometric)............................................ ........ 750 to 1150 mbar a
alternative ranges:
.................................................................. ........ 35 to 1300 mbar a
.................................................................. ........ 35 to 2600 mbar a
.................................................................. ........ 35 to 3500 mbar a
Maximum safe working pressure........................ ......... 4375 mbar a
Display
Read-out................................... .........999999 LCD digits, 13.6mm high
.................. .................. with additional 16 text characters
Overload indication
Error code flashing................. .........at nominal 110% FS
Response............................................ ........ nominal 2 readings/second
Resolution 0.01 mbar (e.g., 1013.25 mbar a)