DPI 740 User Manual 25
K0200 Issue No. 1
3.3 Set-up Menu
(Figure 3-8)
To change other less used facilities and background tasks,
enter the set-up menu by pressing the SET push-button.
Pressing SET again steps to the next sub-menu set. In each
sub-menu the function push-buttons F1 and F2 are used to
select the required task.
Figure 3-8 Set-up Menu
BATTERY
UNITS
MAX/MIN
FILTER
CAL
TIME_OUT
Instrument enters set-up menu
reset MAX/MIN store
MAX/MIN function
set filter window
FILTER function
Carry out a one-point calibration
Carry out a two-point calibration
Set: baud rate, data bits, stop bits,
parity and handshaking
set: time-out period, time in minutes
enable/disable time-out
configure TIME-OUT
Note:
Using the MODE push-button
steps back one menu selection.
configure units of
pressure measurement
battery capacity (%)
COMMS
PIN
PRINT
F1
F2
F1
F2
F1
perform CALIBRATION
F1
enter print period
F1
configure RS232
F2
enter new PIN number
F2
SET
SET
SET
SET
see
processing
measurements