K0200 Issue No. 1
5 0 DP I 7 4 0 U s e r M a n u a l
Set-up commands
Command: SA- Set instrument address
Format: SA = <nn>
Description: Set instrument address to <nn>. Changes the
network address of the instrument in the
range 0 to 98. Address 99 is used for global
commands.
Example: SA=10 set instrument address to 10
Query: SA? query instrument address
Reply: SA=10
Command: SU - Set-up "regular" units
Format: SUn = <units index>
Description: Selects the pressure units that appear as
default selection F2 in measurement mode.
Both required parameters must be specified
i.e. the push-button sequence n (1, 2 or 3) and
the <units index>.
Example: SU1=0 sets first units (1) to mbar
SU2= 18 sets second units (2) to inHg
Query: The corresponding query command returns
the index to the pressure units selected for
the specified option.
SU1?
Reply: SUn=<units index>
SU1=0