DPI 740 User Manual 49
K0200 Issue No. 1
Command: PR - Get reading
Format: PR?
Description: Requests the reading.
Query: This is a query only command.
Example: OR?
Reply: OR = 1000.00
reading in mbar
Command: PA - Enable automatic sending of
reading
Format: PA = k
Description: Enables automatic sending of input reading.
The value `k` corresponds to `send every `k`th
conversion`. If `k` is 0, the automatic send
feature is disabled.
Example: PA = 10
send value every 10th conversion.
PA = 0 disable the automatic feature.
Query: A query responds with an integer of the
automatic send interval in conversions.
Example:PA?
Reply: PA = 10
Command: PM- Resets process maximum and
minimum values
(also see PC command `<` and`>`)
Format: PM
Description: Sets maximum and minimum values equal
to the current reading, in effect resetting both
values.
Query: There is no corresponding query command.