K0200 Issue No. 1
4 8 DP I 7 4 0 U s e r M a n u a l
Example: T tare
PC=T(IR)
Full tare of input reading.
PC=T(IR,100.00)
Define tare of input reading, subtracts 100
from reading. The value specified in the
command is in the currently selected units.
Note
: This tares the value but does not change the displayed
value.
Example: < minimum
PC=<(IR) Defines the minimum value of the input
reading. To reset this value use PM
command.
Example: > maximum
PC=>(IR) Defines the maximum value of the input
reading. To reset this value use PM
command.
Example: Q QFF
PC=Q(IR,200,20)
Define the derived QFF input value changing
both the <height> and <temp> parameters.
Example: A altitude (also see command PA)
PC=A(IR,120.00)
Define the derived ALTITUDE input value changing the
current or default setting of <datum pressure> to 120.00.