K0200 Issue No. 1
2 8 D P I 7 4 0 U s er M a n u a l
Time-out
The time-out period can be set between 1 and 15 minutes
(default 1 minute). In the enabled state, if no push-buttons
have been pressed, the time-out switches the instrument "off"
after the time elapsed. Pressing the ON/OFF push-button "on"
resets the time-out and switches the instrument back on.
When this function is disabled the instrument can only be
switched off by the ON/OFF push-button.
Notes
1. The time-out function is disabled in calibration mode.
2. When switching the instrument on, the time-out can be
disabled by pressing MODE and on/off push-button
together.
Adjusting the Time-out Period
Proceed as follows:
1. Make sure the instrument is in measurement mode and
displaying pressure.
2. Press SET until the status line shows "TIME_OUT".
3. Press F1 to select time-out settings.
4. Use F1 and F2 to adjust to the required time-out
period.
5. Press SET.
6. Press F1 to enable (ON) the time-out function and press
F2 to disable (OFF) the function.
Note
:The instrument stores the adjust time-out period as the
default value in nonvolatile memory.