K0200 Issue No. 1
5 2 DP I 7 4 0 U s e r M a n u a l
Command: CP - Enter Calibration Point
Format: CP = <value>
Description: The calibration point value is given by
<value> in the currently selected units.
Query: A query of this parameter CP? returns the
number of calibration points already
recorded by this command. The instrument
also indicates that it is ready for the next
calibration point.
Reply:
example
CP = 0 no cal points recorded
CP = 2 2 cal points recorded
After entering all the points required to calibrate the
instrument, the accept calibration command CA must be sent
to update the data in the instrument's nonvolatile memory.