K0200 Issue No. 1
5 4 DP I 7 4 0 U s e r M a n u a l
Command: CD - Calibration date
Format: CD = <dd/mm/yy>
Description: Sets calibration date, only valid in calibration
mode.
Example: CD = 24/01/97
set cal date to 24 January 1997.
Query: CD? query date of last calibration.
Reply: CD = 24/01/97
Example of a two point calibration
IU = 0 set-up input as mbar
PP = 123 PIN entered put instrument
into calibration mode
CT = 1 perform two point calibration
CP = 800 applied pressure 800 mbar
CP =1100 applied pressure 1100 mbar
CA calibration accepted
CD calibration date
Command: PP - PIN protection
Format: PP = <pin>
Description: This command is used to change the PIN of
the instrument which is used to protect the
calibration data and configuration.
<pin> = 123 calibration PIN code
Example: PP= 123 PIN entered put instrument into
calibration mode
Query: There is no corresponding query command.