K0200 Issue No. 1
6 0 DP I 7 4 0 U s e r M a n u a l
Table 6-1
Table of Error Definitions
The 16-bit hexadecimal mask defined below can be specified
using the AE command. It is used for automatic error reporting
by the RE message transmitted by the instrument.
Bit Error Description
0 syntax error Set if the command syntax was not understood.
1 parameter error Set if the parameters in the command were out of
range or not valid.
2 configuration The configuration parameters can only be changed
error if the configuration PIN is sent with the commands. If
the PIN is not sent or is incorrect this error is reported.
3 address error An invalid address packet was received.
4 checksum The received command checksum did not match
error with the calculated checksum. With this error, the
command is not executed and this error reported.
5 zero error An error occurred while trying to zero a measurand,
probably because the zero offset is too large.
6 calibration Error in the instrument's calibration procedure;
error not enough calibration points.
7 sequence error A valid command was received, but can not be
processed because the instrument was not in the
correct state to execute command.
8 command not This command was not available on this instrument.
available
9 range error Reading outside range.
10 reserved
11 reserved
12 reserved
13 reserved
14 reserved
15 reserved