Maintenance
This section gives procedures to maintain the unit in a 
good condition. You must return the unit to the supplier for 
all repairs. 
Maintenance - Clean the unit
Clean the case with a moist, lint-free cloth and a weak 
detergent. Do not use solvents or abrasive materials.
Maintenance - Replace the batteries    
To replace the batteries, refer to B1. Then re-attach the 
cover.
Make sure that the time and date are correct. The 
calibration facility uses the date to give service and 
calibration messages.
All the other configuration options stay in memory.
Calibration
Note: GE can provide a calibration service that is traceable 
to international standards.
We recommend that you return the instrument to the 
manufacturer or an approved service agent for 
calibration.
If you use an alternative calibration facility, make sure that 
it uses these standards.
Calibration - Before you start
To do an accurate calibration, you must have:
• the calibration equipment specified in Table 6.
• a stable temperature environment: 70 ± 2°F (21 ± 1°C)
Table 6: Calibration equipment
Before you start the calibration, make sure that the time 
and date on the instrument are correct (Table 1).
Selection sequence: 
➤
 Select Task (Table 2) 
➤
 Set Up (Table 1) 
➤
 Calibration 
➤
 
Table 7: Calibration options
When you select a channel, the display shows the 
applicable instructions to complete the calibration.
When the calibration is complete, select Calibration Due 
and set the new calibration date for the instrument.
Function Calibration equipment
Pressure An applicable pressure standard 
(primary or secondary) with a total 
uncertainty of 0.01% reading or 
better.
mA mA calibrator. 
Accuracy: Refer to Table 9.
B1
1
Display: Enter 
Calibration PIN
23
Menu:
Select Channel
45
▲
▼
+
■ ■
▲
▼
Factory PIN = 4321 (Table 7)
Options
(If applicable)
Description
(* Optional item)
P... To calibrate the specified IDOS pressure input 
(P1 or P2*).
 IDOS ... UMM only. To calibrate the specified IDOS UMM. 
Refer to the user manual for the IDOS UMM.
mA DPI 802/802P only. To calibrate the mA input.
Calibration Due: To set the date of the next 
calibration for the instrument.
After the specified calibration date, there is a 
warning message. There is a selection box to stop 
the warning.
To change the calibration PIN (Personal 
Identification Number).