Table 10: DPI 800 series pressure connections
Table 11: Maximum pressure (+ port)
Table 12: Maximum pressure (Reference port)
Specification - Electrical connectors (A2)
Ranges Pressure connections
(One IDOS range)
Pressure connections
(Two IDOS ranges)
g/d: 30 psi g (2 bar g)
Standard or Premier
1/8 NPT female (+ port) + 1/8 NPT female reference
port OR
G1/8 female (+ port) + G1/8 female reference port
Two 1/8 NPT female (+ port) + 10-32 UNF reference
port OR
Two G1/8 female (+ port) + M5 reference port
g/d: > 30 psi g (2 bar g)
Standard only
1/8 NPT female (+ port) OR
G1/8 female (+ port)
Two 1/8 NPT female (+ port) OR
Two G1/8 female (+ port)
g/d: > 30 psi g (2 bar g)
Premier only
1/8 NPT female (+ port) + 10-32 UNF reference port OR
G1/8 female (+ port) + M5 reference port
Two 1/8 NPT female (+ port) + 10-32 UNF reference
port OR
Two G1/8 female (+ port) + M5 reference port
sg or a: all ranges G1/8 female (+ port) OR
1/8 NPT female (+ port)
Two 1/8 NPT female (+ port) OR
Two G1/8 female (+ port)
Ranges: g/d, sg, a MWP Maximum transient /
intermittent pressure
5 psi (350 mbar) 2 x FS 4 x FS
> 5 psi (350 mbar) 1.2 x FS 2 x FS
Ranges: g/d only MWP
5 psi (350 mbar) 2 x FS
10 to 15 psi
(700 mbar to 1 bar)
1.2 x FS
30 psi (2 bar) 30 psi (2 bar)
Range (Measure) 0 to ±55 mA
Accuracy 0.02% of reading + 3 counts
Temperature coefficient
14 … 50°F, 86 … 122°F
(-1010°C, 30 … 50°C)
0.0011% FS / °F
(0.002% FS / °C)
Switch detection Open and closed. 2 mA current.
Loop power output 24 V ± 10%
HART
®
resistor
250 (menu selection)
Connectors (A2) Three 0.16 in (4 mm) sockets