Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
149
JOB CARD PIST044–001 – Innova 2000 System Interconnect
14
7 of 10
11
5-4 Run the cable Set Group 4
Group 4 links the 3 Monitor suspension to the Technical area.
This cable set of 24 meters is delivered with the suspension.
Suppress, if needed the plastic bag and the connector protections.
Run the cable set in the gullet from the 3 monitor suspension to the technical area.
Keep enough loop to enable the bridge movement.
Set of cables includes 7 cables for ACAB cabinet.
If cable set is too short use the extended cable set
12
delivered following select process.
Extended cable set extends the video lines, the remote control cable and the power distribution.
The extended cable set has a junction box (yellow box
13
or wall plate) to be connected to the
3 monitor suspension cable set, other side is connected to ACAB bulkhead.