Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
31
1 ENVIRONMENTAL REQUIREMENTS/LIMITATIONS
1-1 Room Climate
Avoid extremes in temperatures.
Table 13 – INNOVA 2000 SYSTEM ROOM CLIMATE REQUIREMENTS (BY COMPONENT) – RELATIVE
HUMIDITY AND TEMPERATURE
RELATIVE HUMIDITY
(NON–CONDENSING)
TEMPERATURE
COMPONENT
IN–USE STORAGE IN–USE STORAGE
MIN. MAX MIN. MAX MIN. MAX MIN. MAX
Innova LC
Positioner
5% 90% 5% 85% +5 C +38 C –40 C +70 C
POSTIT
Cabinet
10% 90% 10% 80% +5 C +35 C –40 C +70 C
Omega
Patient table
5% 90% 5% 90% +5 C +35 C –40 C +70 C
Maxiray 150 V5
X–Ray tube assy
N/A N/A N/A N/A +5 C +40 C –20 C +80 C
Innova Digital
Detector (2)
5% 70% 70% +10 C +32 C +10 C +40 C
ACAB
Cabinet
10% 90% 10% 90% +10 C +35 C –20 C +70 C
VMP
Cabinets
30% 80% 95% +16 C +32 C –40 C +80 C
Coolix 2200
Chiller
N/A N/A N/A N/A +13 C +40 C –40 C +80 C
Detector Chiller
HEAT DRY 1
10% 80% 5% 95% 10 C 40 C –40 C
(1)
+70 C
(1)
Innova 2000
Operator console
20% 80% 5% 95% +5 C +40 C –40 C +70 C
DL flat panel
10% 80% 50% 95% +10 C +40 C –25 C +45 C
DL Infrared
emitter/receiver
30% 80% 20% 85% +10 C +35 C –20 C +70 C
DL keypad
8% 90% +10 C +35 C –20 C +70 C
Note: (1) Shipped within GEMS packaging.