Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
79
3 MOUNTING REQUIREMENTS
3-1 Floor Loading and Recommended Mounting Methods
See Table 22. To obtain floor loading and recommended mounting methods for components not
specified in Table 22, refer to the appropriate component Pre-Installation Manual listed in Chapter 2.
Table 22 – INNOVA LC SYSTEM FLOOR LOADING, WEIGHTS, AND MOUNTING METHODS
PRODUCT OR
NET
WEIGHT
DIMENSIONS
LOAD
BEARING
WEIGHT/
OCCUPIED
MOUNTING
PRODUCT
OR
COMPONENT
WEIGHT
KG (LBS)
LENGTH WIDTH HEIGHT
BEARING
AREA
INCHES (MM)
OCCUPIED
AREA
MOUNTING
METHOD
Innova LC
Positioner
700
(1543)
See Illustration 19 and
Illustration 19 b
Circle
diameter
23.62
(600 mm)
Recommended:
Through-Bolts
(12)
Alternates:
On Grade 5/8
in.
Anchors (12)
Above Grade
3/4 in. Anchors
(12)
See Illustration 39
and Illustration 40
(Mounting hard-
ware provided by
GEMS)
Omega Table
590
(1,300)
See
note 1
See Illustration 20
22.5 x 16.9
(571.5 x 429)
492.3 lb/ft
2
(2,410 kg/m
2
)
Same as Innova
LC Positioner
Postit Cabinet
301.5
(664)
See Illustration 22
17.54 x 33
(445x838) 808 kg/m
2
ACAB Cabinet
430
(946)
See Illustration 23
23.63 x
35.44
(600x900) 796 kg/m
2


290
(640)
473
(
18.62
)
555 1980
21.85 x
15 75
1,306 kg/m
2
Recommended
for both cabinets
3/8 in or 10 mm


325
(716)
(18
.
62)
See
note 4
555
(21.85)
1980
(77.95)
15
.
75
(555 x 400)
1,477 kg/m
2
3
/
8 in or 10 mm
(12) anchors to
floor and wall
COOLIX 2200
104
(229)
730
(28.75)
See
note 4
470
(18.5)
686
(27)
Four casters Not applicable