Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
51
Radiation Shield
A counterbalanced radiation shield can be supplied as an option with an Innova 2000 System. The
radiation shield longitudinal axis is not always the same as the OMEGA Patient table longitudinal axis.
A counterbalanced radiation shield can be supplied with a pair of rails. To mount the rails, you will have
to install 3 m steel beams before delivery of the Innova 2000 System.
The 2 steel beams can be used for a DIA product. In this case, the distance of 330 mm (13 inches)
should be changed to adjust both beams to the new product support.
Illustration 18 – CEILING REQUIREMENTS FOR COUNTERBALANCED RADIATION SHIED
2 STEEL BEAMS (3 M or 10 ft LENGTH) MOUNTED TO ADDITIONAL CEILING
STRUCTURE TO BE INSTALLED BEFORE THE DELIVERY OF THE EQUIPMENT
RAD SHIELD ANCHORED WITH 10 SCREWS (10MM DIAM.)
MAXIMUM PULL STRENGTH ON ONE BOLT = 300 N
MOVABLE LOAD 950 N
LONGITUDINAL
AXIS
ISOCENTER
330 (13)
660.4 (26) 660.4 (26) 660.4 (26) 660.4 (26)
710
(Should be adjusted
according to product
provided)
Max 400
(15)
(28)
Cable Support for TV Monitor Cables
A cable support (cable drape) is provided with an Innova LC System.
The cable support kit contains:
1 B2054 EK
1 B2055 ED
Note: In Americas the Cable Support Kit must be provided locally by the Customer
(e.g. CPGE55 from Unistrut).