Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5 pim 2337741–100
83
Illustration 40 – CABLE CONDUIT FOR ON–GRADE FLOOR ANCHOR KIT
Note:
The key hole has been redisigned to permit a 60 mm
(2.5 in.) water conduit for a separate access for water pipes
used with a MAXIRAY 150.
Push 76 (3.0) hole in center.
Reinsert slug and tack well in one spot hole is for 63.5 (2.5)
internal diameter conduit tack well from bottom if possible.