8 Room Light Distribution
Innova 2100-IQ, 3100/3100-IQ and 4100/4100-IQ has the ability to control an external relay that
applies power to the room light (dry contacts).
The relay is to be provided by the hospital or contractor.
The wire size to connect to the C2 cabinet is 1.5 mm
2
(AWG 14).
Relay rating is max 600 VAC
The room light distribution is wired between the C2 cabinet and the PDB.