Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 8 Replacement Procedures 8-5
8-3-1 Loading Base Image Software with DVD
NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the
network and remove all transducers.
NOTE: Please ensure AC adapter is connected during system upgrade!
1.) Insert the disk labeled “System & Application Software” into the DVD ROM drive. Connect the
DVD ROM drive to the system.
2.) Properly turn off the scanner by momentarily pressing the Power On/Off Switch. Select “Full
Maintenance Reboot” from the System Exit menu.
3.) If the system will not shutdown normally, hold down the Power On/Off Switch until the light
turns from green to amber.
4.) Turn on the scanner. System will detect the DVD-RW automatically.
5.) Press any key to continue when below message display as shown in Figure 8-3 on page 8-5 .
Figure 8-2 Shutdown Dialog Box
Figure 8-3 Update message