GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 3 System Setup 3-3
3-1-3 Installation Warnings
1.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing.
NOTE: For information regarding packing labels, refer to LABELS ON PACKAGE.
2.) After being transported, the unit may be very cold or hot. If this is the case, allow the unit to
acclimate before you turn it on. It requires one hour for each 2.5×C increment it's temperature is
below 10×C or above 40×C.
3-1-4 Safety Reminders
Equipment damage possibility. Turning the system on without acclimation after arriving at site
may cause the system to be damaged.
Table 3-3 Time for Settlement
°C
60 55 50 45 40 35 30 25 20 15 10 5 0 -5 -10 -15 -20 -25 -30 -35 -40
°F
140 131 122 113 104 95 86 77 68 59 50 41 32 23 14 5 -4 -13 -22 -31 -40
hrs
864200000002468101214161820
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to
its operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a
three to two prong adapter. This defeats safety grounding.
Do not operate this unit unless all board covers are securely in place.
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the LOGIQ e/LOGIQ e
Vet/LOGIQ i/Vivid e and kept near the unit for quick reference.
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
probe is within FDA limits, avoid unnecessary exposure. Ultrasound energy can produce heat
and mechanical damage