Putting the Device Into Operation and Performance Check
227 490 02-C Marquette Responder® 3000 15
Parameter Comment Factory
Defaults
Options User Setup
ALARM LIMITS
HR Limit 40/160 OFF, 15 to 300
increments of 5
SpO
2
Limit
---/90 OFF, 60 to 99
etCO
2
Limit
---/20 OFF, 5 to 76
increments of 1
ECG
Print on Alarm autom. recorder start on violation of limit
value
off on/off
Lead Fail Alarm audio signal indicating disconnected
electrode
off 30 s/off
Alarm Tone audio signal indicating violation of an
alarm limit
off on/off
QRS Beep off low/middle/high / off
Muscle Filter suppression of motion artifact on on/off
Gain for ECG display 1 cm/mV 0.5; 1; 2 cm/mV
Lead Channel 1 I standard leads,
paddle acquisition
Lead Channel 2 II standard leads, SpO
2
Lead Channel 3 III standard leads, SpO
2
,
etCO
2
DEFIB
Print on shock automatic recorder start on shock on on/off
Operating Mode** choice of the operating mode semiautomatic/button semiautom./button
semiautom./password
semiautom.
manual
Autosequence energy selection 200 J, 200 J, 360 J 150 to 360 J per shock
Pacemaker default pacer rate 60 P/min 30 to 200 P/min
SpO
2
C-LOCK C-Lock ECG synchronization off on/off
SpO
2
Integ. Time SpO
2
integration time 8 s 4 s, 8 s, 12 s
DATE/TIME**
Change clears all existing settings.
Date Format** day.mon.year day.mon.year
mon/day/year
Entry of date and time
DEVICE
Display
screen display (SmartFlip = display flips
automatically when defibrillator is placed
in vehicle mounting unit)
normal normal, reverse,
SmartFlip
Volume valid for all audio signals high high/low/middle
Cont. Printout continuous recording or 14-second strip off on/off
Analysis continuous ECG analysis on on/off
AC Line Filter** elimination of AC line interference 50 Hz off/50 Hz/60 Hz
Language** selection of the language
Factory Default restores factory defaults
User** text or name (20 characters)
PASSWORD**
entry of the password 111 000/999
for config** protects access to configuration menu off on/off
EVENT TEXTS** entry of event texts
BATTERY battery maintenance program
on/off
OPTIONS entry of option code to unlock option
** not affected by reactivation of factory defaults