Error and System Message
74 Marquette ResponderĀ® 3000 227 490 02-C
17 Error and System Messages
Message Effect Explanation Remedy
D-RAM Error device defect, device
cannot be used
notify service
S-RAM Error device defect, device
cannot be used
notify service
FLASH Error device defect, device
cannot be used
notify service
EEPROM Error restricted use for
emergencies only,
defibrillation possible
defective user setup (e.g.
user setup cannot be
saved)
press one of the function keys,
notify service
Charge Energy Error charged energy differs
from selected energy, use
device only in emergen-
cies
error occurred last time
the device was used
(before switched off) and
was saved
press one of the function keys,
trigger test discharge, switch
device off and on again; if error
message recurs, notify service
Time Base Error restricted use for
emergencies only,
defibrillation possible
all time-related data (e.g
heart rate) may be
erroneous
press one of the function keys,
adjust time; if error message
recurs, notify service
Battery battery almost depleted charge battery
HR, etC alarm, violation of alarm
limit
Electrode high ECG or defibrillation
electrode impedance
reapply electrode
Check selector energy selector is not
indexing properly;
defibrillation possible
set energy selector to the exact
position; if error message recurs,
notify service
Energy high internal spoons are
connected and selected
energy is above 50 J;
charged energy higher
than selected energy;
defibrillation possible
select a lower value (internal
spoons) or trigger test discharge;
if message recurs, notify service
Energy low charged energy below
selected energy;
defibrillation possible
(e.g. battery depleted)
trigger test discharge; if message
recurs, notify service
check battery
Device Behavior Explanation Remedy
Device not functioning
properly
software problem switch device off and on again;
notify service