EasyManua.ls Logo

GE Marquette Responder 3000

GE Marquette Responder 3000
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Defibrillation / Non-Synchronized Defibrillation
227 490 02-C Marquette Responder® 3000 21
Shock Counter
The number of delivered shocks is indicated below
the energy value.
This counter is reset to 0 when you set the energy
selector to
.
The number of delivered shocks is also shown on
the recording strip (Figure 5-5, intervention report
d). This counter, however, counts all shocks since
the device was turned on and is reset to 0 only
when the device is turned OFF.
Ending Therapy
Once therapy has ended, set the energy selector
to
for continued monitoring of the patient.
If there is no need to monitor the patient, switch
off the defibrillator by setting the energy se-
lector to "Off".
Clean the paddles and the device as outlined in
chapter 18 "Cleaning, Maintenance".
Defibrillation of Children
Warning
Damage to Myocardium — Please note that chil-
dren require less energy for successful ventricular
defibrillation than adults. For the first defibrilla-
tion shock delivered to babies and small children,
select an energy of approx. 2 J/kg body weight.
For subsequent shocks, the energy may be in-
creased to 4 J/kg body weight.
Risk of Skin Burns — The full electrode must be
in contact with the skin surface (use the small
contact surface of the paddles / pads for children).
The paddles have two different contact surfaces; a
large one (can be removed) for the defibrillation of
adults and a smaller one for the defibrillation of
children.
Remove the large contact surface for pediatric use:
Press on the lock button 1 (Figure 4-5).
Slide the contact surface 2 towards the front
and take it off the paddle.
When re-installing it, the large contact surface
must audibly click into place.
1
2
Figure 4-5. Removing the large contact surface
from the paddles

Related product manuals